>1600KV Range
HIVG 2600KV-390KJ Impulse Voltage Generator
Home > Systems > Impulse Generator > Impulse Generator > >1600KV Range

HIVG 2600KV-390KJ Impulse Voltage Generator

Lead time: 120 days

Mode: HIVG 2600KV-390KJ

Related Standards: 
* IEC 60060-1-2010 High-voltage test techniques - Part 1: General definitions and test requirements
* IEC 60060-2-2010 High-voltage test techniques. Part 2: Measuring systems
* IEC 61083-1-2001 Instruments and software used for measurement in high-voltage impulse tests - Part 1: Requirements for instruments
* IEC 61083-2-1996 Digital recorders for measurements in high-voltage tests - Part 2: Evaluation of software
* ANSI/IEEE 4a-2001 Techniques of High-Voltage Testing
* ANSI/IEEE 376-1975 Measurement of Impulse Strength and Impulse Bandwidth
* ANSI/IEEE 1122-2007 Digit Recorders for Measurements in High-Voltage Impulse Tests
* ASTM D69-D2484 Electrical Insulation ()
* ASTM D2518-Latest Electrical Insulation (
)

Technical Parameters: 
1) Rated Voltage:  ±2600kV
2) Rated stage voltage:  ±200kV
3) Rated charging voltage:  ±100kV
4) Rated energy: 390kJ
5) Stages: 13stages
6) Generator overall capacitance  0.1153uF(pulse capacitor parameter: 100kV 3uF; 26units)
7) Stage energy: 30kJ
8) Impulse voltage waveform parameter:
     a) When object capacitance in the range of 0~3800pF
     b) Standard lightning impulse voltage full-wave
          T1=1.2mS±30%, T2=50mS±20%, oscillation at peak value less than 5% of amplitude
     c) Standard switching impulse voltage wave : 250±20%, 2500mS±60%
     d) Standard Chopping Wave: Adopt chopping device to generate 2~6us standard lighting chopping wave,
         Chopping wave dispersibility deviation <100ns.
     e) This two impulse voltage waveform parameter and deviation meet the requirement of GB311 and GB16927.
9) Minimum output voltage >15%Un
10) Synchronous range: stage voltage is between 20%~100% of rated voltage.
       Positive and negative polarity synchronous range is less than 20%.
11) Synchronous discharge out of control rate <2%
12) Sparking range: 10%-100%Un
13) Charging voltage instability <±1.0%
14) Voltage utilization factor:
       when load less than 1000pF, lightning impulse voltage utilization factor >=90%, switching impulse voltage utilization factor >=75%,
       when load less than 3000pF, the lightning impulse voltage utilization factor >=85%, switching impulse voltage utilization factor >=65%
15) Working time:
       When voltage above than 2/3 rated voltage, the system can continuous working each 120s (charging and discharging).
       When the voltage is below 2/3 of rated voltage, the system can continuous working each 60s (charging and discharging)
16) Generator height: 9800mm
17) Generator weight: 14500kg

Principle Diagram: 

Where To Buy

Related Researches