≤800kV Range
HIVG 300KV-20KJ Impulse Voltage Generator
Home > Systems > Impulse Generator > Impulse Generator > ≤800kV Range

HIVG 300KV-20KJ Impulse Voltage Generator

Lead time: 60 days

Mode: HIVG 300KV-20KJ

Related Standards: 
* IEC 60060-1-2010 High-voltage test techniques - Part 1: General definitions and test requirements
* IEC 60060-2-2010 High-voltage test techniques. Part 2: Measuring systems
* IEC 61083-1-2001 Instruments and software used for measurement in high-voltage impulse tests - Part 1: Requirements for instruments
* IEC 61083-2-1996 Digital recorders for measurements in high-voltage tests - Part 2: Evaluation of software
* ANSI/IEEE 4a-2001 Techniques of High-Voltage Testing
* ANSI/IEEE 376-1975 Measurement of Impulse Strength and Impulse Bandwidth
* ANSI/IEEE 1122-2007 Digit Recorders for Measurements in High-Voltage Impulse Tests
* ASTM D69-D2484 Electrical Insulation (Ⅰ)
* ASTM D2518-Latest Electrical Insulation (Ⅱ)

Technical Parameters: 
1) Rated Voltage:300kV
2) Rated stage voltage:100kV
3) Rated charging voltage:100kV
4) Rated energy:20kJ 
5) Each stage impulse capacitor: 1.334μF/100kV 
6) Rated impulse capacity: 0.445μF
7) Stages:3 stages 
8) Stage energy:6.67kJ
9) Impulse voltage waveform parameter: 
A. Standard lightning impulse voltage full-wave: 1.2±30%us /50±20%us,
Amplitude: <3%, oscillation at peak value less than 5% of amplitude
B. Standard lightning chopping wave 2~5us 
C. This two impulse voltage waveform parameter and deviation meet the requirement of GB311 and GB16927.
10) Minimum output voltage efficiency:
Standard lightning impulse voltage full-wave≥90% (No test object)
11) Synchronous range: stage voltage is between 10%~100% of rated voltage.
Positive and negative polarity synchronous range: less than 20%.
12) Synchronous discharge out of control rate <2% (Within the range: 10%~100% of rated voltage )
13) Sparking range: 10%-100%Un
14) Minimum output voltage: ≥10% rated voltage.
15) Charging voltage instability <±1.0%
16) Deviation between charging voltage and reference voltage  <±1.0%
17) Working time: 
When voltage above than 70% rated voltage, the systems can continuous working each 120s (charging and discharging). 
When the voltage is below 70% of rated voltage, the system the system can continuous working each 60s (charging and discharging).
18) Under Maximum charging voltage, minimum discharge time not less than 60s.


Principle Diagram: 