Mode: HCRS 450kV/15750kVA
Rated voltage: 450kV Rated current: 35A
Rated input voltage: 3.3kV Rated frequency: 60Hz
System noise level: <72dB @ rated voltage and current
System partial discharge level<=2pC@450kV
System voltage measuring accuracy: <±1% (According to IEC30060-1, measuring in 10% -100% rated voltage. ) System quality factor Q>=40 (no water termination and resistance load)/ Q>=10 (with water termination and resistance load)
System insulation level: withstand 2min @110% of rated voltage
System working time: one hour on, one hour off, 4cycle per day.
System load capability:
@450kV, two reactor serial connection, maximum load <=206.3nF (Include divider and coupling capacitor, 60Hz)
@225kv, two reactor parallel connection, maximum load <=825.2nF (Include divider and coupling capacitor, 60Hz)
@225kV, single reactor, maximum load <=412.6nF (Include divider and coupling capacitor, 60Hz)
12. System waveform deviation: <=1%
Mode: HCRS 225kV /7875kVA
Cylinder typeSingle phase Frequency: 60Hz
Indoor use Rated Capacity: 7875kVA
Rated Voltage: 225kV (Two reactor serial connection voltage is 450kV)
Rated Current: 35A Cooling methods: ONAN
Waveform deviation ≤1% Noise: ≤75dB
Single reactor rated inductance: 17H
Two reactor serial connection rated inductance: 34.1H
Quality factor Q>=65
Over Current Ability: Continuous working for 300S under 110% rated current (38.5A), then thermal damage or distortion of the winding shall not be occurred.
Over Voltage Ability: Overvoltage at 110% UH (495kV) for 120s., then it shall not happened any insulation damage for the testing transformer.
Working time and temperature: One hour on and One hour off under rated load, everyday 4cycle.Winding temperature raise not surpass 65K, oil temperature raise less than 55K
Linearity: 10%-100% reactance, error≤±1%
System overall structure requirement
1. When two module serial connection, its must guarantee:
1.1 Transmission mechanism should be connected convenient and synchronism, the transmission should be flexible and low noise.
1.2 Each module’s location and fixation should be convenient, reliable, firstly, fix reactor, and then connect transmission shaft;
1.3 Its should be convenient for the current leading circuit between modules.
1.4 Each module voltage distribution almost correspond with each other.
1.5 The connection of HV lead wire should be fixed and convenient;
2. Each reactor Z=f(δ) (The relations between impedance and air gap) characteristic curve should be almost same as each other, at the same air gap position, the different between impedance not surpass 5% of minimum value.
3. Reactor should has reliable air gap limit switch, and equip with air gap sensor, the user can read the iron coil’s gap on the control console.
4. Transmission device should has back-up protection treatment, in case of damage to the transmission from phase sequence opposite or air gap limit switch dysfunction
5. The system’s tuning and resonance function should be excellent. The system should not detuning when regulating device raise voltage, resonant(adjusting air gap) raise voltage or under any test voltage (not surpass rated voltage). And, when the exciter transformer output a relative low voltage (1-1.5kV), the system can also tuning to resonant state.
6. Regulating device should has the function regulation from zero position and auto return-to-zero function
7. The reactor’s air gap will back to the middle position after testing.
8. Regulator has the protection function when motor out of control.
9. Dimension requirement on transportation and crane
- Internal structure:
①. 30-Q-130 Grain-Oriented Electrical Steel,
②. There are earthing baffle between winding and iron coil, the earthing is reliable.
③. We do many improment on the system; abandon some unreasonable structure and creat our new resonant. New variable inductance power frequency resonant reactor’s working life not less than 60 years.
④. The iron steel has been done the chip removing treatment. And then form a entirety with curing agent, which lower the noise; equip guide rail between adjustable iron coil and fixed iron coil, to make sure the iron coil move vertical or horizontal. We have been leader of the power frequency variable resonant system, moreover, we cooperate with famous university and research institute and improve ours quality continuously.
Principle Diagram:
HIMALAYAL provides innovative products that comprise high voltage test equipment, full-set high voltage test instruments, on-site test systems and on-line monitoring systems, and services regarding tests, diagnostics and asset monitoring for clients in power industry. We help to ensure safety and reliability in the power generation, transmission and distribution.